Unstable and High Angular Readings in Laser Shaft Alignment: Causes and Solutions (Field Case)

Short summary: If the angular misalignment value on a laser shaft alignment tool changes from one measurement to the next while the parallel offset stays constant, the problem is usually in the setup rather than in the tool itself. In this article we share the symptoms, likely causes and a step-by-step checklist, based on a real case we encountered in the field.

The case: “The tool is unstable and reads too high”

A customer abroad contacted us while trying to align a pump–motor set with an SKF TKSA 41 laser shaft alignment tool. The videos they sent showed the following:

  • Even though the machine was not touched, the angular misalignment value was different in every measurement.
  • The parallel offset, on the other hand, stayed almost constant.
  • The foot (shim) values calculated from the angle varied several times over, and in the horizontal correction the tool kept saying “left”, then “right”.
  • A measurement taken on the same machine with another laser alignment tool was stable and showed that the machine was already within tolerance.

The customer naturally suspected the tool. After reviewing the videos, we suggested checking the setup instead of looking for a fault. A short while later we heard back: once the distance between the measuring units was brought into the operating range stated in the manual and entered correctly, the tool started working normally. No replacement or service was needed.

Diagram showing the distance between the S and M measuring units in laser shaft alignment
The M unit is on the movable machine and the S unit is on the stationary machine. The distance between them is the basic input of the calculation.

Why does the angle fluctuate while the parallel offset does not?

Laser alignment systems calculate angular misalignment roughly as follows:

Angular misalignment ≈ difference between two measurement points ÷ distance between the measuring units

For this reason, if the distance entered into the tool is smaller than the actual distance, or if the units are positioned outside the tool's operating range, the same small difference turns into a much larger angle. The slightest vibration during the measurement is magnified in the same proportion, and the result changes with every measurement. Parallel offset is much less affected by this distance, so it keeps looking stable. The pattern we saw in the videos was exactly this.

There is one more consequence: because the tool also calculates the foot values from this angle and the entered distances, a distance error translates directly into a wrong shim recommendation.

Likely causes of an unstable angular reading

From the most common to the least common:

  1. The distance between the measuring units is entered incorrectly or is outside the operating range. The distance is entered by hand. Entering an estimate instead of measuring it with a tape measure, or mounting the units too close to or too far from each other, is the most common mistake.
  2. Mechanical play. A loose rod, a chain-type mounting bracket or the unit locking knob can also make only the angle fluctuate. The units must not be held by hand during measurement, and they must not be used as handles to turn the shaft.
  3. The units are swapped. The S unit must be on the stationary machine and the M unit on the movable (to be adjusted) machine. A reversed setup reverses the correction directions.
  4. External factors. Vibration from nearby machines, air currents rising from hot surfaces or strong direct light can affect the readings.
  5. Tool failure. This is the least likely cause and can be separated from the others with simple tests (below).
SKF TKSA 41 S measuring unit (the unit mounted on the stationary machine)
S unit: mounted on the stationary machine (e.g. the pump).
SKF TKSA 41 M measuring unit (the unit mounted on the movable machine)
M unit: mounted on the movable machine to be adjusted (e.g. the motor).

Step-by-step checklist

  1. Measure the distance with a tape measure and compare it with the value in the tool. Make sure the distance between the units is within the operating range stated in the user manual.
  2. Verify the position of the units: S on the stationary machine, M on the movable machine.
  3. Tighten all connections: rods, chains, unit knobs. Do not touch the units during measurement.
  4. Watch the live detector screen. The measuring unit information screen of the tool shows the raw detector values live. If these values fluctuate noticeably while the machine is stationary, the cause is vibration, play or the tool.
  5. Run a single-shaft test. Mount both units on the same rigid shaft or pipe and measure. The result should be close to zero and repeatable. If it is not, the tool should be sent for service.
  6. Repeat the measurement three times in a row. Results taken without touching the machine should be close to each other.
  7. Check the date of the calibration certificate.

Important: do not add shims until the measurement is stable

Adding shims based on an unstable measurement can throw off a machine that is actually within tolerance. Do not adjust the feet until the values are repeatable. For shim selection and soft foot checks, see our machinery shim and soft foot guide.

SKF TKSA 41 laser shaft alignment tool mounted on the shafts
SKF TKSA 41: measuring units on the shafts, display unit in front.

Frequently asked questions

If the angular value fluctuates but the parallel value is constant, is the tool faulty?
Most of the time, no. This pattern is most often caused by an incorrectly entered distance between the measuring units or by mechanical play. The live detector screen and the single-shaft test let you tell whether the tool itself is faulty.

If two different alignment tools give different results, which one should I trust?
First check the setup of both tools: the distances, the position of the units and the tightness of the connections. The measurement with the correct setup that gives repeatable results should be taken as the reference.

What should the distance between the measuring units be?
Every tool has an operating range stated in its user manual. The distance must be within this range, and the actual value measured with a tape measure must be entered into the tool.